Sensor and Controller Catalog
Lab Services
Quality Certification Lab
Thermocouple Calibration Watlow offers testing for application temperatures other than the standard points in a range from -100 to 2300°F (-74 to 1260°C), depending on material. Use outside of the temperature limits of ASTM E230 is not recommended.
A lab service technician uses a metallurgical microscope to examine the microstructure of the metallic components in Watlow’s mineral insulated metal sheathed cable, XACTPAK.
Thermocouple Calibration Table
Sensor Type T Thermocouple J Thermocouple E Thermocouple K or N Thermocouple
Temperature 32 to 700°F (0 to 371°C) 32 to 1400°F (0 to 760°C) 32 to 1600°F (0 to 871°C) 32 to 2300°F (0 to 1260°C) -110 to 1200°F (-80 to 650°C)
Specifications
ASTM E220/ASTM E230
ASTM E220/ASTM E230
ASTM E220/ASTM E230
ASTM E220/ASTM E230
PT 385 100Ω RTD
ASTM E644
Notes: • Thermocouple calibrations to ASTM E207 possible (contact factory).
Recommended Sensor Length for Calibration
Temperature -100 to 500°F (-74 to 260°C) 500 to 2000°F (260 to 1093°C) 2000 to 2300°F (1093 to 1260°C)
Minimum Length
Maximum Length
6 in. (150 mm)
60 in. (1525 mm)
18 in. (450 mm)
10 ft (3 m)
20 in. (500 mm)
10 ft (3 m)
Notes: Longer sensors may be calibrated if they are in coil form.
Recommended RTD Length for Calibration Temperature
Minimum Length
Maximum Length
-100 to 500°F (-74 to 260°C) 500 to 1200°F (260 to 650°C)
6 in. (150 mm)
60 in. (1525 mm)
18 in. (450 mm)
10 ft (3 m)